TF L | [Transfer Function - Generate CTF, in doc file] |
TF | [Transfer Function - Generate image showing effect of defocus on CTF] |
TF CT | [Transfer Function - Generate a binary, phase flipping, complex, CTF correction image] |
TF CTS | [Transfer Function - CTF correction with SNR, image/volume] |
.SPHERICAL ABBERATION CS [mm]: 2.0
[Enter the
spherical aberration coefficient.]
.DEFOCUS [A], ELECTRON VOLTAGE [Kev]: 20000, 300
[Enter the amount of
defocus,
in Angstroms. Positive values correspond to underfocus
(the preferred region); negative values correspond to overfocus.
Next, enter the energy of the electrons in Kev.
(Note: operation still accepts the legacy input of electron wavelength
lambda [A] instead of voltage)].
.NUMBER OF SPATIAL FREQUENCY POINTS: 128
[Enter the length of the 1D array. This is the size of the square image that
will be CTF corrected.]
.PIXEL SIZE [A]: 2.8
[Enter the
number of Angstroms per pixel
in the digitized micrograph.]
.SOURCE SIZE [1/A], DEFOCUS SPREAD [A]: 0.005, 0
[Enter the
size of the illumination source
in reciprocal Angstroms. This is the size of the source as it appears in
the back focal plane of the objective lens. A small value
results in high coherence; a large value, low coherence.
Enter the estimated magnitude of the defocus
spread
corresponding to energy spread and lens current fluctuations.]
.AMPL CONTRAST RATIO [0-1], GAUSSIAN ENVELOPE HALFWIDTH: 0.1, 0.15
[Enter the ACR and the
GEH.
The Gaussian envelope parameter specifies
the 2 sigma level of the Gaussian (see note 2 for details).]
NOTES